Monitoring of radiation dose to medical staff during 125I seeds loading
10.3760/cma.j.issn.0254-5098.2015.07.013
- VernacularTitle:125I粒子植入前装载过程中医护人员的剂量监测
- Author:
Yao YUE
;
Haisheng LIU
;
Junjie WANG
;
Chengxiang GUO
;
Yuliang JIANG
;
Haitao SUN
- Publication Type:Journal Article
- Keywords:
125I seeds;
Thermoluminescence dosimeter;
Dose surveillance
- From:
Chinese Journal of Radiological Medicine and Protection
2015;35(7):536-538
- CountryChina
- Language:Chinese
-
Abstract:
Objective To analyze the dose of radiation to medical staff during 125I seeds loading Methods The radiation dose at different distances was measured by using thermoluminescence dosimeters (TLD),and the safe range for medical person was calculated.The doses and annual accumualive doses to the medical staff were estimated.Results The dose attenuation rates were 77.61%,98.04%,98.79%,99.30%,99.71% and 100% at distance of 10,20,30,40,50 and 100 cm from seeds tank,respectively.The dose to fingertips,chest,lens and thyroid of doctor were 51.08,35.50,34.73 and 33.78 μGy,and the annual dose reached 12.77,8.88,8.68 and 8.45 mGy when the number of annual operations was 250.The attenuation rate was 79.60% and 28.36% of inside and outside lead glass,respectively.Conclusions The radiation is undetectable at 100 cm away from seeds tank in the process of loading of 125I seeds,and lead glass is necessary for radiation protection.