Intelligent optimum tube voltage technology at dual-source CT coronary artery angiography
10.3760/cma.j.issn.1005-1201.2015.04.012
- VernacularTitle:智能最佳管电压扫描技术在双源CT冠状动脉成像中的应用价值
- Author:
Jun ZHANG
;
Dan HAN
;
Bo HE
;
Xiaojie XIE
- Publication Type:Journal Article
- Keywords:
Tomography,X-ray computed;
Coronary angiography;
Radiation dosage
- From:
Chinese Journal of Radiology
2015;49(4):288-292
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the value of intelligent optimum tube voltage technology at dual-source CT coronary angiography.Methods One hundred and fifty patients with normal body mass index (<25 kg/m2) with clinically suspected coronary heart disease requiring coronary artery CTA examination were collected.According to the different scanning method,patients were randomly divided into 5 groups,30 cases of each group were tested by coronary artery CTA examination.The intelligent optimum tube vohage scanning technology was set to Semi,the tube voltage of A,B,C group were 120,100 and 80 kV (Ref 120 kV,400 mAs),respectively.D and E group were set to on,the reference voltage of group D was 120 kV,and reference current was 400 mAs.The reference voltage of group E was 100 kV,reference current was 400 mAs.Using the 15 section improved segment method for evaluation of coronary artery (lumen diameter> 2 mm and no severe calcification).They were assessed by subjective evaluation (image quality score) and objective evaluation (including vascular CT value,signal-to-noise ratio and contrast to noise ratio).Analyses of the differences between groups were compared with image quality,objective evaluation index and radiation dose by single factor variance.Results The image quality scores in patients scanned with five groups were (3.42±0.63),(3.41±0.54),(3.49±0.33),(3.45±0.43) and (3.48±0.81),there was no statistical difference between the five groups (F=0.634,P=0.105).Each image quality was good,all could clearly show thewhole range of vessels.The CT values of the coronary artery with above groups respectively were (486±82),(554±71),(742±90),(506±81),(561±81) HU.The image noises were (25±12),(32±12),(46±14),(28± 11),(34±12) HU (F=3.430 and 4.332,P<0.05).And the SNR and CNR was no statistic difference between the five groups (P>0.05).The effective radiation dose of five groups are (9.3± 1.3),(6.3± 1.4),(2.8±1.5),(5.7 ± 1.2),(3.9 ± 1.3) mSv,which the difference was statistically significant (F=2.332,P<0.05).Conclusions Intelligent optimum tube voltage technology (Ref.100 kV,400 mAs) at dual-source CT coronary angiography may be feasible in patients with normal BMI.To ensure the quality of the image at the same time,it can effectively reduce the radiation dose.