Clinical evaluation of FW-Ⅱ axial blood pump short-term assistance for treating acute left heart failure
10.3760/cma.j.issn.1001-4497.2014.10.008
- VernacularTitle:FW-Ⅱ轴流泵短期辅助治疗急性左心衰初步临床评价
- Author:
Shengshou HU
;
Hansong SUN
;
Lihuan LI
;
Zujun CHEN
;
Li SHI
;
Yan ZHANG
;
Haibo CHEN
- Publication Type:Journal Article
- Keywords:
Axial blood pump;
Acute heart failure;
Hemodynamics;
Myocardial injury
- From:
Chinese Journal of Thoracic and Cardiovascular Surgery
2014;30(10):599-602
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the safety and efficacy of FW-Ⅱ axial blood pump short term assistance for treating acute heart failure.Methods We selected 5 patients who were difficult to remove cardiopulmonary bypass,and implanted FW-Ⅱ axial blood pump by left atrium-pump-femoral artery.Perioperative hemodynamics,circulating markers of myocardial injury,vW factor levels in plasma and white blood cell-platelet aggregates in different rotational speed were observed and analyzed.Results All patients were assisted for(24.0 ± 2.6) h by FW-Ⅱ axial blood pump with speed of 7 000-9 000 r/min,flow of 1.9-3.0 L/min,and no mechanical dysfunction.Compared with mean artery pressure(MAP),cardiac index(CI) and systemic vascular resistance index (SVRI) before implantation of FW-Ⅱ axial pump was (50.29 ± 6.98) mmHg (1 mmHg=0.133kPa),(1.70±0.23)L· min-1 · m-2 and (2 009.86 ±129.46) dyn· s · cm-5,MAP and CI significantly increased,while SVRI significantly reduced at 8 000 r/min after pump assistance [(65.43 ± 6.90) mmHg,(2.53 ± 0.27) L · min-1 · m-2,(1 578.14 ± 356.70) dyn · s · cm-5,P < 0.01).CK-MB and cTnI levels reached significant decrease after 12 hours of pump assistance and maintained low level [(66 ± 11) IU/L and (8.4 ± 3.8) μg/L,P < 0.01].Compared with 7 000 r/min,vW factor levels significantly increased at 8 500 r/min [(2.59 ± 0.57) U/L vs (1.26 ± 0.43) U/L,P <0.01].Platelet activation and white blood cell-platelet aggregation was the lowest at 8 000 r/min,and reached the most high level at 7 000 and 9 000 r/min [(15 ± 3) %,(33 ± 3) % and (31 ± 5) %)].Conclusion FW-Ⅱ axial flow pump can be effectively used for short-term treatment of acute ischemic left ventricular failure,8 000 r/min is optimum speed to balance hemodynamics and biocompatility.