Influence of laser positioning system to position repeatability in the use of improved belly board
10.3760/cma.j.issn.1006-9801.2014.09.012
- VernacularTitle:激光定位系统应用在改进型腹板中提高放疗患者体位重复性的研究
- Author:
Fei CAO
;
Ming LI
;
Kaiyue CHU
- Publication Type:Journal Article
- Keywords:
Laser positioning system;
Improved belly board;
Setup error
- From:
Cancer Research and Clinic
2014;26(9):624-625
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate influence of laser positioning system to position repeatability of patients treated with radiotherapy for pelvic tumors.Methods Forty patients with pelvic tumors were divided into two groups,using laser positioning system positioning (20 cases) and conventional positioning (20 cases),respectively,CBCT was used to verify and analyze the setup errors of two groups.Results The setup errors on X,Y and Z axes in laser positioning system positioning group were (2.63±1.07) mm,(2.38±1.28) mm and (2.46±1.21) mm,they were (3.04±1.22) mm,(2.55±1.35) mm and (3.09±1.37) mm in conventional positioning group.It has significance reduce on X and Z axes (P =0.040,0.003,respectively) by using laser positioning system,and has no significance on Y axis (P =0.432).Conclusion It can improve patient positioning repeatability effectively by using laser positioning system for patient positioning.