Analysis of the consistency of the detection results of glycosylated hemoglobin (HbA1c) by SMART POCT and hitachi 7180 biochemical analyzers
10.3969/j.issn.1673-4130.2015.12.010
- VernacularTitle:SM ART POCT 与日立7180生化仪检测糖化血红蛋白的一致性分析
- Author:
Jianhong TIAN
;
Yi WANG
;
Jianghua JIA
;
Yusheng ZHANG
- Publication Type:Journal Article
- Keywords:
biochemical analyzer;
glycosylated hemoglobin;
correlation coefficient
- From:
International Journal of Laboratory Medicine
2015;(12):1663-1664
- CountryChina
- Language:Chinese
-
Abstract:
Objective To analyze the consistency of the detection results of glycosylated hemoglobin (HbA1c) by SMART POCT and Hitachi 7180 biochemical analyzers .Methods Firstly the accuracy of HbA1c detected by the SMART POCT and Hita‐chi 7180 biochemical analyzers were evaluated .And then 20 randomly selected samples were tested by both equipments .A relevance analysis between detected data from two equipments was conducted .Results There was a good correlation between the results of the HbA1c detected by two detecting systems ,with the correlation coefficient of r2 =0 .981 1 .Conclusion There is a good consis‐tency in detection results of glycosylated hemoglobin (HbA1c) detected by SM ART POCT and Hitachi 7180 biochemical analyzer .