Experimental study of the monitoring bias of pressure in intubation balloon using handheld pressure gauge
10.3760/cma.j.issn.2095-4352.2014.05.013
- VernacularTitle:气囊压力表间断测量气囊压力值偏差的实验研究
- Author:
Yanxia LIN
;
Dan LIN
;
Biqun CHEN
;
Chao JI
;
Congli YUAN
;
Baochun WANG
- Publication Type:Journal Article
- Keywords:
Management of the cuff pressure;
Monitoring intermittently;
Bias;
Experimental study
- From:
Chinese Critical Care Medicine
2014;26(5):347-350
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the bias between the real pressure and the measured values when handheld pressure gauge (HPG) was used to monitor intermittently the pressure in the intubation balloon,so as to provide some measures for the correct use of HPG.Methods In the first part of the study,HPG was used to measure the pressure with the balloon connected with a three-way tube with which to control the inflation and deflation in a laboratory to measure the pressure in the air bag.After gaining the deviation in this in vitro experiment,it was tested and verified in vivo in adult patients undergoing endotracheal intubation.Results After 132 times of measurements,it was found that measurement with a HPG might result in an inherent loss (3.928 ± 0.291) cmH2O (1 cmH2O=0.098 kPa,t =155.273,P =0.000) between inflation value [(30.000 ± 0.000) cmH2O] and measured value [(26.072 ± 0.291) cmH2O].In addition,after 214 times repeated measurements,the pressure loss during disconnection of the gauge was as high as (1.196 ± 0.954) cmH2O (t=18.348,P=0.000) between filled values [(30.000 ± 0.000) cmH2O] and measured values [(28.804 ± 0.954) cmH2O] and it was named as error loss.At last,the total error was verified by clinical test,and it was (5.270 ± 2.583) cmH2O (t=29.632,P=0.000) between pressure of filled value [(30.000 ± 0.000) cmH2O] and measured value [(24.730 ± 2.583) cmH2O].Conclusions When the balloon pressure was Monitored intermittently with HPG,the real value should be the measured value plus the error.In addition,subglottic aspiration should be done before the connection of the balloon to the gauge to prevent the secretions on the cuff falls into the deeper airway,and to maintain the cuff pressure at 30 cmH2O.