Precision and uncertainty evaluation and comparison analysis of 6 items blood lipids index on different Beckman AU biochemical testing systems
10.3969/j.issn.1673-4130.2015.14.023
- VernacularTitle:血脂6项在Beckman AU 系列多个检测系统精密度、不确定度的评估及可比性分析
- Author:
Jialin XIANG
;
Xiaoli YANG
;
Yan YANG
;
Xuhong OUYANG
;
Niwei HAN
;
Ling YIN
- Publication Type:Journal Article
- Keywords:
precision;
uncertainty;
comparison;
biochemical analysis;
blood lipids
- From:
International Journal of Laboratory Medicine
2015;(14):2007-2008
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the precision and uncertainty and comparison analysis of 6 items blood lipids index among 5 different Beckman AU biochemical testing systems ,such as triglyceride(TG) ,total cholesterol(TCHO) ,high density lipoprotein cholesterol(HDL‐C) ,low density lipoprotein cholesterol(LDL‐C) ,apolipoproteins A1(APOA1) ,apolipoproteins B(APOB) .Meth‐ods According to the document the EP15‐A2 of national Committee for Clinical Laboratory Standards ,6 items blood lipids index were respectively detected by 5 different Beckman AU biochemical testing systems to obtain precision and comparison .The intra‐and inter‐precision and results of comparability among different system were low than 1/4 or 1/3 or 1/2 CLIA′88 as evaluation standard ,respectively .The measurement uncertainty of these items were evaluated by the calibrator uncertainty and internal quality control and external quality control .Results The intra‐and inter‐or day‐to‐day precision and relative bias were accepted by clinical requirements .The expanded measurement uncertainty for TG was 0 .079 mmol/L and 0 .035mmol/L .The expanded measurement uncertainty for TCHO was 0 .248 mmol/L and 0 .157 mmol/L .The expanded measurement uncertainty for HDL‐C was 0 .144 mmol/L and 0 .018 mmol/L .The expanded measurement uncertainty for LDL‐C was 0 .140 mmol/L and 0 .186 mmol/L .The ex‐panded measurement uncertainty for APOA1 was 0 .148 mmol/L and 0 .090 mmol/L .The expanded measurement uncertainty for APOB was 0 .104 mmol/L and 0 .058 mmol/L .Conclusion The results of 6 items blood lipids index respectively show well preci‐sion and significantly correlation among 5 different Beckman AU biochemical testing systems and the results were comparable ,and the influence factor of detection results were expression directly by evaluating the measurement uncertainty of 6 items blood lipids index .This way of assessment is simple .