Quality assessment of seven kinds of sero-enzyme tests in China
10.3760/cma.j.issn.1009-9158.2015.05.007
- VernacularTitle:我国七个血清酶检验项目检测质量分析
- Author:
Jing WANG
;
Jie ZENG
;
Ying YAN
;
Haijian ZHAO
;
Rong MA
;
Jiangtao ZHANG
;
Weiyan ZHOU
;
Tianjiao ZHANG
;
Cuihua HU
;
Donghuan WANG
;
Chuanbao ZHANG
- Publication Type:Journal Article
- Keywords:
Diagnosis tests,outine;
Laboratory proficiency testing;
Quality assurance,health care
- From:
Chinese Journal of Laboratory Medicine
2015;(5):313-317
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the analytical quality of different analytical systems in measuring seven kinds of sero-enzymes consisting of Alanine Aminotransferase(ALT), Aspartate Aminotransferase (AST),γ-Glutamyltransferase(GGT), Lactate Dehydrogenase(LDH), Creatine Kinase(CK), α-Amylase (AMY) and Alkaline Phosphatase(ALP).Methods Data from 2013 routine chemistry external quality assessment (EQA) and Enzymes Trueness Verification(ETV) were collected.1 450 and 165 participating laboratories were selected respectively for investigation.Analytical systems of participating laboratories were classified into 6 kinds,i.e.imported matching system(AI), domestic matching system(AH), systems consisting of imported reagents and corresponding calibrators(BI), systems consisting of domestic reagents and corresponding calibrators ( BH ) , unmatched systems using imported calibrators ( CI ) and unmatched systems using domestic calibrators ( CH ) .Total error, bias and coefficient of variation within laboratories ( CVI) were calculated from the data of 2013EQA and ETV The proportion of laboratories meeting the desirable and the optimal criteria derived from biology variation were analyzedby EXCEL2010 with coincidence rate (CR) above 85% as evaluation criterion.Results The AI and CI occupied more than 70%among six systems, CH occupied approximate 15% and the other systems were less than 10%.The
range of the average of ETV′s total errors , EQA′s total errors, absolute value of bias and CVI of seven kinds of sero-enzymes were 6.2%-27.8%, 4.0%-7.0%, 4.2%-25.1% and 3.6%-4.6% respectively. Accuracy, bias and within-laboratory imprecision were judged by CR of ETV′s total errors, ETV′s bias, CVI and EQA′s total errors respectively and comparability between different systems was evaluated.It turned out that the results of analytical systems of enzymes except ALP were comparable, the accuracy of systems of enzymes except AMY, ALP and GGT, LDH of AI, the within-laboratory imprecision of enzymes except LDH, AMY, ALP and AST of AI, CH could meet the desirable criteria.The bias of all systems of seven kinds of sero-enzymes were undesirable.Conclusions The analytical quality of routine testing of seven kinds of sero-enzymes could fulfill the clinical requirement generally in China.