Influence of platform switching versus platform matching on the marginal bone resorption around im-plant:A Meta-analysis
10.3969/j.issn.1001-3733.2015.03.018
- VernacularTitle:种植体边缘骨吸收:平台转移与平台匹配的 Meta 分析
- Author:
Zihuan SUN
;
Rong XIA
;
Jiliang XU
;
Xi MIN
;
Chun LIU
- Publication Type:Journal Article
- Keywords:
Dental implants;
Platform switching;
Marginal bone resorption;
Meta-analysis
- From:
Journal of Practical Stomatology
2015;(3):378-383
- CountryChina
- Language:Chinese
-
Abstract:
Objective:To evaluate the effects of platform switching and platform matching system on the marginal bone resorption a-round implant.Methods:Randomized controlled trials (RCTs)that compared marginal bone loss around platform-switched implants with platform matched prostheses were selected from PubMed,EMbase,CBM,CNKI and other electronic databases supplemented by hand search and retrospective collection of literature published or unpublished between 1 991 -201 4.The literature based on inclusion and exclusion criteria was screened by 2 revieweres independently,the quality of the included studies was evaluated,the data were extracted using RevMan 5.2 software for Meta-analysis.Results:1 4 studies with 1 331 implants were included.Meta-analysis showed that peri-implant bone resorption in the platform switching group was significantly less than that in the platform matching group[MD =-0.51 ,95% CI:(-0.72 -0.30),P <0.01 ].Subgroup analysis showed that the implant-abutment diameter difference >0.45 mm (unilateral)was more favorable to implant marginal bone preservation.Conclusion:The present data suggest that platform-switched technology is more conducive to implant bone preservation than platform-matched method.