The commutability of reference materials for serum glucose measurements
10.3760/cma.j.issn.1009-9158.2015.05.003
- VernacularTitle:血清葡萄糖参考物质在24种常规分析系统上的互通性研究
- Author:
Tianjiao ZHANG
;
Jie ZENG
;
Mo WANG
;
Chuanbao ZHANG
;
Jiangtao ZHANG
;
Haijian ZHAO
;
Jing WANG
;
Donghuan WANG
;
Weiyan ZHOU
;
Rong MA
;
Cuihua HU
;
Wenxiang CHEN
- Publication Type:Journal Article
- Keywords:
Blood glucose;
Reference standards;
Reference values
- From:
Chinese Journal of Laboratory Medicine
2015;(5):296-300
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the commutability of certified reference materials, external quality assessment program materials and calibrators for serum glucose measurements which were performed in 24 routine measurement procedures.Methods 35 fresh patient specimens and some reference materials were analyzed by isotope dilution liquid chromatography tandem mass spectrometry ( as the comparative method) and 24 routine measurement procedures (as the evaluated methods).The relationships between the results from the evaluated method and the comparative methods were evaluated to identify the commutability.Results It showed that 5 certified reference materials, 2 trueness verification materials, and 5 calibrators were commutable in all 24 routine measurement procedures.The other samples were displayed the presence of commutability issue in different degrees.Conclusion It is important to pay more attention to the problems brought by commutability of reference materials in clinical laboratory.