Performance verification for Olympus Au5400 in automatic biochemistry analyzer
10.3969/j.issn.1673-4130.2015.04.028
- VernacularTitle:OlympusAU5400全自动生化分析仪光路系统性能验证
- Author:
Yuelin YANG
;
Dachun HU
;
Lichun WANG
- Publication Type:Journal Article
- Keywords:
automatic biochemistry analyzer;
optical system;
performance verification;
medical standard
- From:
International Journal of Laboratory Medicine
2015;(4):498-500
- CountryChina
- Language:Chinese
-
Abstract:
Objective To verify and confirm the optical performance of the Olympus AU5400 Automatic Analyzer is satisfied for the professional requirements.Methods According to the Medical Standard of the People's Republic of China:Automatic chemistry analyzer(YY/T 0654-2008),the Analyzer Modular Ⅰand Ⅱ optical performance were measured to verify the stray light,the ab-sorption accuracy,linear range,stability and repeatability.Results The stray light absorption of tow modular was 4.789 4-5.158 9;The absorption accuracy of potassium dichromate standard solution (1-1)was (-0.002 1--0.024 7),and solution (1-2)was (-0.008 6--0.067 7);The D-value of stability that the highest absorption of standard solution (1-1)minus the lowest was 0.000 5-0.001 5;The CV% of repeatability for standard solution (1-2)was 0.02%-0.04%;Therof absorption linear was 0.999 6-0.999 9. Conclusion The optical performance of Olympus AU5400 Analyzer conforms to the standard of YY/T 0654-2008.