Performance evaluation of AU2700 biochemical analyzer for detecting serum glycated albumin
10.3969/j.issn.1673-4130.2015.05.016
- VernacularTitle:AU2700生化分析仪检测血清糖化清蛋白的性能评价
- Author:
Huihui YAN
;
Qinfeng HUANG
;
Pingping AN
;
Jian LIAO
;
Min CHEN
- Publication Type:Journal Article
- Keywords:
glycated albumin;
diabetic;
performance evaluation
- From:
International Journal of Laboratory Medicine
2015;(5):613-614
- CountryChina
- Language:Chinese
-
Abstract:
Objective To use the AU2700 biochemical analyzer to evaluate the performance of the reagent kit for determining se-rum glycated albumin(GA)by the enzymatic method.Methods On the basis of the relevant documents of CLSI-EP,the perform-ance evaluation of the Lucica GA-L reagent kit for detecting GA was performed on the precision,accuracy,linearity range and refer-ence range;the correlation between the results detected by the two kinds of reagent kit (enzymatic method)was analyzed.Results The reagent had high precision(CV<3%);the dilution had good linearity;the average recovery rate of GA was 97.7%,the average recovery rate of albumin(ALB)was 96.43%;the detection results by the two kinds of reagent kit had good correlation (r2 =0.976, P <0.01).Conclusion The Lucica GA-L reagent kit shows the good analytical performance on the AU2700 biochemical analyzer, which is consistent to the manufacturers statement and can be better used in clinic.