Comparison of the determination results of potassium by blood-gas analyzer and dry chemical analyzer
10.3969/j.issn.1673-4130.2014.21.039
- VernacularTitle:血气分析仪与干化学仪测定血钾离子结果比较
- Author:
Xiaoping MU
;
Chunmin LU
;
Ziling LIU
;
Cunwei JI
;
Cai ZHOU
- Publication Type:Journal Article
- Keywords:
blood-gas analyzer;
dry chemical analyzer;
potassium;
method comparison
- From:
International Journal of Laboratory Medicine
2014;(21):2951-2952
- CountryChina
- Language:Chinese
-
Abstract:
Objective To analyze the comparability of potassium results detected by blood-gas analyzer and dry chemical analy-zer .Methods Dry chemical detection system was used as comparison method (X) ,and blood gas analyzer systems was used as the experimental method (Y ) .Arterial blood samples collected from ICU newborns were detected respectively by the two methods .If SE of medical decision level was less than or equal to 1/2 TEa decided by CLIA′88 ,the results was acceptable .Results There was a linear correlation in potassium concentrations of the two detection systems(r=0 .976 ,P<0 .01) .But among three medical decision levels ,the SE of the two detection systems only acceptable at 3 .0 mmol/L .Conclusion The potassium concentration of blood gas analyzer is lower than that of dry chemistry analyzer .The potassium concentration of dry chemistry analyzer should be taken as a reference to diagnose and clinical treatment .