The research on the comparability of thyroid hormone test results using two different detection systems
10.3969/j.issn.1673-4130.2014.21.040
- VernacularTitle:两套不同检测系统甲状腺激素测定结果的可比性分析
- Author:
Jianqiong WANG
;
Hua NIU
;
Ling CHEN
;
Qin ZHANG
;
Zhuxian PING
- Publication Type:Journal Article
- Keywords:
thyroid hormone;
detection system;
comparability
- From:
International Journal of Laboratory Medicine
2014;(21):2953-2954
- CountryChina
- Language:Chinese
-
Abstract:
Objective To analysis of the thyroid hormone test results of Abbott Architect ci16200 automatic biochemical-im-mune analyzer (Abbott system) and Roche Cobas e601 automated immunoassay analyzer (Roche system) .Methods The thyroid hormone(including T4 ,FT4 ,T3 ,FT3 ,TSH ) levels of 93 cases of serum samples were respectively detected by Abbott system (CMIA) and Roche system (ECLIA) ,and the results were analyzed .Results The thyroid hormone test results of the two systems had a good correlation ,and the coefficients of T4 ,FT4 ,T3 ,FT3 ,and TSH were 0 .960 ,0 .962 ,0 .976 ,0 .900 ,and 0 .999(P<0 .01) . However ,there were significant difference of the detection results of the two systems(P<0 .05) .Conclusion The thyroid hormone test results of Abbott system and Roche system are not comparable .