Comparison of glycosylated hemoglobin levels detected by 3 kinds of analytic instruments
10.3969/j.issn.1673-4130.2014.19.044
- VernacularTitle:3种仪器检测糖化血红蛋白结果分析
- Author:
Weijun ZHANG
;
Shengqi HUANG
;
Lianmei LUO
;
Xiujuan LI
;
Yanfen ZHENG
- Publication Type:Journal Article
- Keywords:
glycosylated hemoglobin;
ion exchange chromatography;
affinity chromatography
- From:
International Journal of Laboratory Medicine
2014;(19):2672-2673
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare levels of glycosylated hemoglobin detected by 3 kinds of analytic instruments .Methods 75 samples were measured by D-10 glycosylated hemoglobin automatic analyzer(ionexchange chromatography) ,HA-8160 glycosylated hemoglobin automatic analyzer(affinity chromatography) and 7170A automatic analyzer(immune turbidimetry) .Results were tested by the homogeneity of variance ,the one-way analysis of variance and correlation analysis .Results There was a positive correlation between D-10 glycosylated hemoglobin automatic analyzer and HA-8160 glycosylated hemoglobin automatic analyzer(r2 =0 .996) , the linear equation was Y=0 .953X+0 .519 .There was a positive correlation between D-10 glycosylated hemoglobin automatic ana-lyzer and 7170A automatic analyzer(r2 =0 .996) ,the linear equation was Y=0 .925X+0 .576 .There was a positive correlation be-tween HA-8160 glycosylated hemoglobin automatic analyzer and 7170A automatic analyzer(r2 =0 .998) ,the linear equation was Y=0 .969 X+0 .081 .Conclusion In the premise of quality control in laboratory ,three different instrument can use to detect the level of glycosylated hemoglobin .