Determination of Meropenem in Urine by Fluorescence Resonance Energy Transfer Spectroscopy
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:荧光共振能量转移光谱法测定尿液中美罗培南的含量
- Author:
	        		
		        		
		        		
			        		Qiulian LIU
			        		
			        		;
		        		
		        		
		        		
			        		Songqing LI
			        		
			        		
		        		
		        		
		        		
 
			        		
			        		
		        		 
- Publication Type:Journal Article
- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Meropenem;
			        		
			        		
			        		
				        		Eosin-Y;
			        		
			        		
			        		
				        		Fluorescein;
			        		
			        		
			        		
				        		Urine;
			        		
			        		
			        		
				        		Fluorescence resonance energy transfer spectroscopy
			        		
			        		
	        			
        			
        		
- From:
	            		
	            			China Pharmacist
	            		
	            		 2014;(12):2016-2018
	            	
            	
- CountryChina
- Language:Chinese
- 
		        	Abstract:
			       	
			       		
				        
				        	Objective: To establish the determination method for meropenem in urine by fluorescence resonance energy transfer spectroscopy. Methods:An LS-55 fluorescence spectrophotometer was used. The solution of meropenem, fluorescein, eosin Y, BR buffer and cetyl trimethyl ammonium bromide was respectively added into a 10 ml colorimetric tube. At λex of 455 nm and λem of 547 nm, the fluorescence intensity F and F0 of the system and the reagent blank sample was detected. Using ΔF = F-F0 as signal re-sponse value for meropenem, the content was calculated. Results: The linear regression equation of meropenem in urine was as fol-lows:ΔF=33. 8C+53. 4 (r=0. 991 7), the linear range was 0. 5-10 μg·ml-1 with the detection limit of 0. 13 μg·ml-1, the re-covery of 98. 9%-103. 0% and RSD of 0. 3%-0. 4%. Conclusion:The method is rapid and accurate, which can be used in the phar-macokinetic study of meropenem.