The Usefulness of the PCL Power Calculation Computer Program: 'POWER!'.
- Author:
Jong Mo SEO
1
;
Jin Hak LEE
Author Information
1. Department of Ophthalmology, Seoul National University College of Medicine, Korea. jjhlee@plaza.snu.ac.kr
- Publication Type:Original Article
- Keywords:
IOL power calculation;
Postoperative refractive value
- MeSH:
Humans
- From:Journal of the Korean Ophthalmological Society
2002;43(1):23-28
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
PURPOSE: To assess the usefulness and reliability of the IOL power calculation computer program ('POWER!'). METHODS: Calculation times and projected postoperative refractive values of 100 patients with 'POWER!' program were compared with those of the program mounted on the Humphrey R A/B scan system (model 820) by SRK-II and SRK/T formula. RESULTS: There was no significant difference between the projected postoperative refractive values of 'POWER!' program and Humphrey R A/B scan system (paired-t test, p>0.05). Calculation time of 'POWER!' was shorter than that of the program mounted on the Humphrey R A/B scan system incalculating more than 3 sets of IOL (paired-t test, P<0.05) and there was no internal calculation error (ztest, p>0.05). CONCLUSIONS: 'POWER!' is a useful and reliable program for IOL power calculation using SRK-II and SRK/T formula.