Application analysis of XT-4000i blood cell analyzer in body fluid cell count
10.3969/j.issn.1673-4130.2014.23.045
- VernacularTitle:XT-4000i血细胞分析仪在体液细胞计数的应用分析
- Author:
Junjie CHEN
;
Rui SHI
;
Mei XU
;
Fei CHEN
- Publication Type:Journal Article
- Keywords:
XT-4000i blood cell analyzer;
body fluids cell count;
white blood cell;
red blood cell
- From:
International Journal of Laboratory Medicine
2014;(23):3255-3256
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the application value of the XT-4000i blood cell analyzer in body fluid cell count.Methods 113 cases of body fluid specimen were collected in the hospitalized patients from September 2013 to February 2014.Then RBC and WBC counts in the collected specimens were detected whitin 1 h by XT-4000i blood cell analyzer and the manual detection method, the RBC count values were divided into 3 levels:L1 100~1 000 ×106/L,L2 1 001 ×106 ~100 000 ×106/L and L3 > 100 000 × 106/L;WBC count values were divided into 2 levels:L1 1~50×106/L and L2 >50×106/L.The correlation between the two kinds of test methods was analyzed.Results The results of RBC and WBC counts detected by the XT-4000i blood cell analyzer and the manual method had a higher correlation.The correlation coefficients were 0.931,0.996,0.865,0.942 and 0.988.Conclusion The XT-4000i blood cell analyzer can be applied in clinical fluids cell count.