Comparison and evaluation of the determination of serum CEA between high-throughput ELISA and ECLIA
10.3969/j.issn.1673-4130.2014.22.037
- VernacularTitle:高通量ELISA法与电化学发光法测定血清癌胚抗原的比较
- Author:
Xiaofeng MU
;
Aifeng ZHOU
;
Ziyun ZHAO
;
Juan CHEN
;
Hua YU
- Publication Type:Journal Article
- Keywords:
enzyme linked immunosorbent assay;
carcinoembryonic antigen;
electrochemiluminescence
- From:
International Journal of Laboratory Medicine
2014;(22):3088-3089
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare the performance of high‐throughput ELISA and ECLIA in the determination of carcinoembry‐onic antigen (CEA) .Methods The CEA concentration of serum samples were respectively determinate by high‐throughput ELISA and ECLIA ,and the results were compared .Results Two kinds of detection methods could both accurately reflect the concentration of serum CEA .There was no significant difference between the results of two methods (P> 0 .05) .Within the linear range ,the CEA result of high‐throughput was correlate closely with that of ECLIA (r=0 .922 8 ,P<0 .01) .Conclusion High‐throughput ELISA can accurately detect the serum CEA concentration .