Contamination status of dental unit waterlines in three general hospitals
10.3969/j.issn.1671-9638.2014.12.005
- VernacularTitle:3所综合医院口腔科综合治疗台水路污染状况调查
- Author:
Xiaoling ZHENG
;
Yuwen ZHONG
;
Huizhen CHEN
;
Yajing WANG
;
Bingshu WANG
;
Chunhua HAN
;
Xiuting SHEN
;
Qin ZOU
- Publication Type:Journal Article
- Keywords:
dental unit waterline;
bacterial contamination;
handpiece spray water;
flushing water;
source water;
water quality monitor;
healthcare-associated infection
- From:
Chinese Journal of Infection Control
2014;(12):720-723
- CountryChina
- Language:Chinese
-
Abstract:
Objective To realize the contamination status of dental unit waterlines (DUWL)in general hospitals, and provide scientific evidence for making preventive measures.Methods Three hospitals were selected for study, water source adopted by hospital A,B and C was running water,reservoir water,and filtered water through reverse osmosis filtration system respectively,specimens of dental handpiece spray water and flushing water of dental chair units were collected quarterly,total bacterial colony in water were detected.Results The qualified rate of source wa-ter,handpiece spray water,and flushing water in hospital A was 75.00%(3/4),0 (0/40)and 0 (0/40)respectively,col-ony count of handpiece spray water and flushing water was (1.20×103 -5.53×104 )CFU/mL(M=3.80×104 CFU/mL) and (2.11×104 -1.66×105 )CFU/mL(M=4.80×104 CFU/mL)respectively.The qualified rate of source water,hand-piece spray water,and flushing water in hospital B was 50.00%(2/4),60.00%(24/40)and 72.50%(29/40)respectively, colony count of handpiece spray water and flushing water was (0.00 -3.71 ×106 )CFU/mL(M=83.00 CFU/mL)and (0.00-2.39×106 )CFU/mL(M=72.00 CFU/mL)respectively.The qualified rate of source water,handpiece spray wa-ter,and flushing water in hospital C was 100.00%(4/4),55.00%(22/40)and 65.00%(26/40)respectively,colony count of handpiece spray water and flushing water was (0.00-6.20×103 )CFU/mL(M=96.00 CFU/mL)and(0.00-1.63×103 )CFU/mL(M=87.50 CFU/mL)respectively.Conclusion Water of DUWL in general hospitals is seriously con-taminated,disinfection and standardized management of source water and DUWL must be strengthened.