Performance verification of precision and accuracy of Olympus AU5421 detection system
10.3969/j.issn.1673-4130.2014.24.049
- VernacularTitle:OlympusAU5421检测系统精密度和正确度性能验证
- Author:
Jing QIAN
;
Dachun HU
;
Liqiong YANG
;
Yuelin YANG
;
Xiufang LI
- Publication Type:Journal Article
- Keywords:
performance verification;
precision;
accuracy
- From:
International Journal of Laboratory Medicine
2014;(24):3415-3416
- CountryChina
- Language:Chinese
-
Abstract:
Objective To analyze the precision and accuracy of the Olympus AU5421 automatic biochemical analyzer in order to verify the performance of this detection system declared by the manufacturer.Methods The Precision and Accuracy of User Au-thentication-Guide for Approval Second Edition(CLSI EP15-A2)was used to perform the routine detection on the Olympus AU5421 automatic biochemical analyzer.The systematic precision and accuracy were analyzed.Results Under this experiment condition,the precision and accuracy of the Olympus AU5421 automatic biochemical analyzer was consistent with the performance declared by the manufacturer.Conclusion The Olympus AU5421 automatic biochemical analyzer system has the high precision and good accuracy, and can be better applied in the clinical routine detection.