Comparative study on application of diluted serum and EXCEL and MVS in linear validation
10.3969/j.issn.1673-4130.2014.16.044
- VernacularTitle:稀释血清及 EXCEL 与 MVS 在线性验证中的应用比较研究
- Author:
Xueping ZHANG
;
Guiying YANG
;
Gang DU
;
Haifeng XU
;
Ping ZHANG
;
Xia GE
- Publication Type:Journal Article
- Keywords:
diluted serum;
linear performance verification;
laboratory techniques and procedures
- From:
International Journal of Laboratory Medicine
2014;(16):2222-2223
- CountryChina
- Language:Chinese
-
Abstract:
Objective To study the application value of diluted serum and EXcelland MVS in the linear performance verifica-tion.Methods The series of concentration samples were prepared by mixing the collected low value mixed serum and the low value mixed serum diluted with 2-3 times of deionized water and the same high concentration serum for conducting the linear experi-ments.The linearity of the experimental results was statistically judged by the two softwares of EXcelland MVS.Results The 3 diluted serum experiments were linear by the judgement of the two softwares;in 6 experiments,the original serum experimental re-sults of CREA were linear by the judgement of EXcelland nonlinear by the judgement of MVS,but the decisive concentration lev-els,the differences of the two regression equations were -2.02 μmol/L and 4.23 μmol/L respectively,which were less than 1/9 of CLIA′88 maximum allowable error under the two concentrations(26.52 μmol/L and 39.78 μmol/L),the remaining 5 experiments were linear by the judgement of the two softwares.Conclusion The low value serum appropriately diluted by deionized water does not affect the linear performance verification;in the linear performance verification,EXcellcan meet the verification needs and MVS is more strict in judging linearity.