The commutability and assignment of target values of proficiency testing/external quality assessment sample in clinical laboratory
10.3760/cma.j.issn.1009-9158.2014.08.018
- VernacularTitle:临床检验能力验证/室间质量评价标本的互换性及靶值的设定
- Author:
Yaling XIAO
;
Haijian ZHAO
;
Wei WANG
;
Zhiguo WANG
- Publication Type:Journal Article
- Keywords:
Clinical competence;
Loboratory techniques and procedures;
Quality control
- From:
Chinese Journal of Laboratory Medicine
2014;37(8):634-636
- CountryChina
- Language:Chinese
-
Abstract:
Key factors for interpreting proficiency testing (PT)/external quality assessment (EQA) results are knowledge of the commutability of the samples used and the process used for target value assignment.A commutable PT/EQA sample,which would be seen as patients' samples,has the agreement of results among different measurement procedures and demonstrates the same numeric relationship.The value assignment for the sample can be done by using of a reference measurement procedure.Its results can be used to assess the accuracy against a reference measurement procedure or a designated comparison method.Noncommutable PT/EQA samples frequently have a matrix-related bias of unknown magnitude,and can't give meaningful information about the relationship of results for patients' samples among different measurement procedures.Its results must be compared to the peer group mean or the median of results from participants who use measurement procedures that are expected to have the same or very similar matrixrelated bias.