Verification of dosimetric methodology for auditing radiotherapy quality under non-reference condition in Hubei province
10.3760/cma.j.issn.0254-5098.2014.06.016
- VernacularTitle:湖北省核查放射治疗非参考条件剂量学参数方法的验证
- Author:
Xinxing MA
;
Suming LUO
;
Zhijian HE
;
Wenshan ZHOU
- Publication Type:Journal Article
- Keywords:
Non-reference condition;
Finger ionization chamber;
Photon beams;
Electron beams;
Absorbed dose
- From:
Chinese Journal of Radiological Medicine and Protection
2014;34(6):461-465
- CountryChina
- Language:Chinese
-
Abstract:
Objective To verify the reliability of TLD-based quality audit for radiotherapy dosimetry of medical electron accelerator in non-reference condition by monitoring the dose variations from electron beams with different field sizes and 45° wedge and the dose variations from photon beams with different field sizes and source-skin distance.Methods Both TLDs and finger ionization chambers were placed at a depth of 10 cm in water to measure the absorbed dose from photon beams,and also placed at the depth of maximum dose from electron beams under non-reference condition.TLDs were then mailed to National Institute for Radiological Protection,China CDC for further measurement.Results Among the 70 measuring points for photon beams,58 points showed the results with a relative error less than ± 7.0% (IAEA's acceptable deviation:± 7.0%) between TLDs and finger ionization chambers measurements,and the percentage of qualified point numbers was 82.8%.After corrected by P,value,62 points were qualified and the percentage was up to 88.6%.All of the measuring points for electron beams,with the total number of 24,presented a relative error within ± 5.0% (IAEA's acceptable deviation:± 5.0%) between TLDs and finger ioization cylindrical chambers measurements.Conclusions TLD-based quality audit is convenient for determining radiotherapy dosimetric parameters of electron beams in non-reference condition and can improve the accuracy of the measuring parameters in connection with finger chambers.For electron beams of 5 MeV < E0 < 10 MeV,the absorbed dose parameters measured by finger ionization chambers,combined with TLD audit,can help obtain the precise and reliable results.