Corrosion resistance of 3M nickel-titanium arch wires and Damon copper-nickel-titanium arch wires in different artificial saliva with different fluoride and pH
10.3969/j.issn.2095-4344.2014.16.008
- VernacularTitle:氟离子和酸对3M镍钛弓丝和Damon铜镍钛弓丝耐腐蚀性的影响
- Author:
Yihui XIONG
;
Yunan LI
;
Xiaoshuang ZHANG
- Publication Type:Journal Article
- Keywords:
corrosion;
orthodontic wires;
saliva,artificial
- From:
Chinese Journal of Tissue Engineering Research
2014;(16):2506-2511
- CountryChina
- Language:Chinese
-
Abstract:
BACKGROUND:In the fluorine-containing acidic environment, it is unclear whether Damon copper- nickel-titanium arch wire can have a corrosion resistance similar to conventional nickel-titanium arch wire and whether copper ion wil affect its corrosion resistance. OBJECTIVE: To evaluate the corrosion resistance of 3M nickel-titanium arch wires and Damon copper-nickel-titanium arch wires in artificial saliva with different fluoride and pH. METHODS:Potentiodynamic polarization curves of 3M nickel-titanium arch wires and Damon copper-nickel-titanium arch wires were measured in four kinds of artificial saliva (pH=7; pH=7, 0.2% fluoride; pH=7, 0.5% fluoride; pH=5, 0.5% fluoride). Then, the corrosion potential, corrosion current density and polarization resistance were acquired, and scanning electron microscope was used for morphological observation after corrosion. RESULTS AND CONCLUSION: In neutral (pH=7) artificial saliva, 0.2% fluoride ions added could shift up the Damon specimen polarization curve, increase the corrosion cerrent density and reduce the polarization resistance (P < 0.05); under the scanning electron microscope, corrosion was observable on the specimen surface. But 3M specimen polarization curve did not move, corrosion current density and polarization resistance substantialy unchanged (P > 0.05), and no significant corrosion was produced. When fluoride concentration was increased to 0.5%, the polarization curves of two kinds of the specimens moved, the corrosion cerrent density increased, the polarization resistance decreased (P < 0.05), and there was significant presence of corrosion on the specimen surface. When pH=5, corrosion phenomenon was more pronounced. These findings suggest that low concentrations fluoride cannot affect corrosion resistance of 3M arch wire, but can reduce the corrosion resistance of Damon arch wire; high concentrations of fluoride and acid can both reduce their corrosion resistance. In addition, the corrosion resistance of 3M arch is stronger than that of Damon arch wire.