The role of 3 external quality assessment programs in assessment of analytical quality on determination of serum creatinine and urea
10.3760/cma.j.issn.1009-9158.2014.05.010
- VernacularTitle:三种外部质量评价计划对实验室血清肌酐和尿素分析质量的作用
- Author:
Chuanbao ZHANG
;
Haijian ZHAO
;
Rong MA
;
Tianjiao ZHANG
;
Jie ZENG
;
Zhiguo WANG
- Publication Type:Journal Article
- Keywords:
Laboratories,hospital;
Quality control;
Creatinine;
Urea;
Uncertainty
- From:
Chinese Journal of Laboratory Medicine
2014;(5):371-375
- CountryChina
- Language:Chinese
-
Abstract:
Objective To describe and compare the roles of 3 external quality assessment programs in assessment of analytical quality of serum creatinine and urea.Methods Research in quality management methods.Sixty-five laboratories those enrolled in the Natonal Center for Clinical Laboratories′programs of routine chemistry external quality assessment ( EQA ) , trueness verification ( TV ) for small molecular metabolites and external comparison of internal quality control ( IQC) simultaneously in 2013 were selected , the performances of those laboratories of serum creatinine (crea) and urea in terms of total errors(TE), bias and CV were obtained by using the above 3 programs, and these performance were assessed against the criterion listed in the analytical quality specifications for routine analysis in clinical biochemistry ( WS/T 403-2012).The failure ratio of 65 laboratories on each performance was calculated , the sensitivity of 3 external quality assessment programs in detection of analytical quality deficiency among clinical laboratories were compared.Results Only 1 laboratory failed in the 1st routine chemistry EQA in terms of TE of creatinine , failure ratio is 1.5%(1/65).Three laboratories failed in the 2nd EQA and caused a failure ratio of 4.6%(3/65).For serum urea, 3 laboratories failed in the 1st routine chemistry EQA with a failure ratio of 4.6%(3/65).Two laboratories failed in the 2nd EQA with a failure ratio of 3.1%(2/65).The failure ratios of creatinine determination in two samples in TV were 41.5%(24/65) and 21.5%(14/65) respectively, and the failure ratio of urea determination were 53.8%( 36/65 ) and 32.3%( 21/65 ) respectively.In the program of external comparison of IQC , the CVs of creatinine and urea determination ranged from 0.7% to 6.2%and from 1.0%to 7.2%respectively, their respective failure ratio range were 15.4%(10/65) and 40.0%(26/65).The failure ratio in routine EQA were much less than those in the other two programs , the laboratories failed in routine EQA program were all failed in trueness verification or /and the comparison of IQC programs, but not vice versa.Conclusions By participating in the programs of routine EQA , TV and comparison of IQC laboratories could assess the performances of inaccuracy , bias and imprecision.Laboratories should participate in different external quality assurance programs to detect their quality issues and get improved.