Experimental study of bioabsorbable iron-based pulmonary artery stent
10.3760/cma.j.issn.2095-428X.2014.06.017
- VernacularTitle:生物可吸收铁基肺动脉支架的动物实验研究
- Author:
Dongpo LIANG
;
Zhiwei ZHANG
;
Shushui WANG
;
Xu ZHANG
;
Zhaofeng XIE
;
Guohong ZENG
- Publication Type:Journal Article
- Keywords:
Bioabsorbable stent;
Iron;
Pulmonary artery;
Congenital heart disease;
Child
- From:
Chinese Journal of Applied Clinical Pediatrics
2014;29(6):459-463
- CountryChina
- Language:Chinese
-
Abstract:
Objective To monitor the degradation and bio-safety features of the bioabsorbable iron-based stent planted in the pig pulmonary artery.Methods A total of 19 miniature pigs were randomly assigned into 3 groups.Three cases were assigned as the control group; ten cases were assigned as stent implantation group A with one stent implanted in the pig pulmonary artery; the other six cases were assigned as stent implantation group B with two stents implanted in the pig pulmonary artery.Follow-up studies for 24 months were conducted in all cases.Repeated measures ANOVA were used to collect and analyze statistical processing data at multiple time points of the follow-up.Results After surgery,pulmonary artery blood flow velocity and right ventricular pressure were measured by echocardiography and cardiac catheterization.There was no significant difference in pulmonary artery blood flow velocity (Fgroup×time =1.06,P =0.40) and right ventricular pressure (Fgroup ×time =0.58,P =0.86) among each group,and no vascular renarrow circumstance was indicated.Pulmonary artery diameter expansion rate between stent group and control group was statistically different (Fgroup =3.57,P =0.05 ; Ftime =12.89,P < 0.001 ; Fgroup ×time =2.99,P =0.03),suggesting that the stent could maintain a good expansion of the role of vascular in the follow-up period even though in degradation.Difference in the level of serum iron concentrations among the groups was not statistically significant (Fgroup ×time =0.94,P =0.52),suggesting that there was no evidence of iron overload.All animals survived to the follow-up endpoint,and no serious side effects caused by stent implantation were found.Conclusions Bioabsorbable iron-based stent planted in the pig pulmonary artery may experience a certain degree of degradation,and it is safe and stabile in animals.