The plasma amino acids biological reference intervals of L-8900 amino acid analyzer
10.3969/j.issn.1673-4130.2014.08.047
- VernacularTitle:L-8900氨基酸分析仪血浆氨基酸生物参考区间的建立
- Author:
Zhijuan FAN
;
Fengmei ZHANG
;
Jie LIU
;
Shuye LIU
- Publication Type:Journal Article
- Keywords:
L-8900 amino acid analyzer;
domestic reagents;
biological reference interval of amino acids;
liver function
- From:
International Journal of Laboratory Medicine
2014;(8):1045-1047
- CountryChina
- Language:Chinese
-
Abstract:
Objective To calculate the biological reference of plasma amino acid on L-8900 amino acid analyzer and to provide reference for clinical diagnosis with domestic reagents replacing original reagent .Methods By testing the original standards and the same batch of plasma (50 cases) ,we compared original reagents with domestic reagents for their performance (including resolution , repeatability ,accuracy ) .We tested the plasma free amino acids of 400 cases of healthy people using the domestic reagents to estab-lish biological reference interval of plasma amino acid ,and do correlation analysis between the amino acids level and liver function . Results (1)Domestic reagents showed high accuracy in the results of 5 consecutive detection of amino acids were high peak separa-tion and high peak retention time and high peak area .(2)Statistics derived biological reference interval of 19 amino acids and 10 kinds of amino acids had significant differences .(3)Correlation analysis showed that ALT and liver function were negatively corre-lated with threonine(P<0 .05) .GLU and valine ,isoleucine ,leucine were positively correlated(P<0 .05) .CHO and negatively cor-related with isoleucine(P<0 .05) .Conclusion Domestic agents can replace the original reagents ,on the basis ,the biological refer-ence intervals of plasma amino acids have great importance to clinical diagnosis and prognosis .