Analysis of electron beam output factors by Monte Carlo method
10.3760/cma.j.issn.1004-4221.2014.03.023
- VernacularTitle:基于蒙特卡洛方法研究电子线输出因子影响因素
- Author:
Zifeng CHI
;
Dan LIU
;
Yong ZHANG
;
Runxiao LI
;
Zhonghao JING
;
Feng FENG
;
Chun HAN
- Publication Type:Journal Article
- Keywords:
Electronbeam irradiation;
Output factor;
Monte Carlo method
- From:
Chinese Journal of Radiation Oncology
2014;23(3):265-268
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the application of the Monte Carlo dose calculation of output factors for electron beams in radiotherapy.Methods The code EGS4/MCTP was used to simulate the head of a medical linear accelerator (Varian 23EX) to calculate the output factors for 6 MeV,9 MeV and 18 MeV electron beams.The source-to-surface distance used was 100 cm.The field sizes ranged from 2 cm × 2 cm to 25 cm × 25 cm.The calculated output factors agreed with the corresponding measured factors which were measured by the IBA Phantom system to within 2%.Then,the output factors of direct articles and indirect articles which were under different energy and various cone-insert combinations were calculated by the code EGS4/MCTP.Results The calculated output factors agreement with the measurements is found to be mostly under the 1% level.The variation of output factors depends on the characteristics of the beams and the modifications that the various cone-insert combinations introduce to these characteristics.Conclusions Monte Carlo dose calculations for electron beams in homogeneous water phantoms have been demonstrated to be accurate under the 1% level in comparison with measurements.The output factors were influenced by energy and cone-insert combinations in complex ways.