Clinical application of IMSure QA software
10.3760/cma.j.issn.0254-5098.2014.04.014
- VernacularTitle:IMSure QA的临床应用研究
- Author:
Jiyong ZHANG
;
Jiayang LU
;
Lili WU
;
Zhijian CHEN
- Publication Type:Journal Article
- Keywords:
Intensity modulated radiotherapy;
Dose verification;
IMSure QA;
MatriXX
- From:
Chinese Journal of Radiological Medicine and Protection
2014;34(4):292-294
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare IMSure QA and MatriXX for intensity modulated radiotherapy plans,and to investigate the credibility of IMSure QA software.Methods Ten patients receiving intensity modulated radiotherapy were randomly chosen for the quality assurance plans with IMSure QA and MatriXX.Results The overall average of percentage pass points in 3% and 3 mm were (98.1 ± 0.8) % with IMSure QA,and (97.9 ±0.6)% with MatriXX(t =0.86,P >0.05).Conclusion IMSure QA can be a reliable verification tool for IMRT QA.