Research on the uncertainty analysis and evaluation of experimental results about sterilization equipment
10.3969/J.ISSN.1672-8270.2014.01.009
- VernacularTitle:医用灭菌设备实验结果的不确定度分析和评定
- Author:
Li CUI
- Publication Type:Journal Article
- Keywords:
Sterilizer;
Calibration specifications;
Uncertainty
- From:
China Medical Equipment
2014;(1):25-27
- CountryChina
- Language:Chinese
-
Abstract:
Objective:It is verified whether JJF1308-2011 standard is scientific and reasonable, whether it is easy to operate and achieve the technical requirements of the testing medical sterilizer temperature by analyzing and evaluation the uncertainty of experimental. Methods:The temperature detection device is used to obtain experimental data in the condition of no-load. The result is concluded after analysis and evaluation of uncertainty. Results:The specification is verified to reach the feasibility and rationality through the no-load test and analysis and evaluation of uncertainty of the results. Conclusion:It is proved that the specification is easy to operate through the experiment. It can meet the requirement of testing the quality and safety of sterilizer.