Experimental determination of dosimetry parameters for Sinko 125I seed source using thermoluminescent dosimeter
10.3760/cma.j.issn.1004-4221.2014.02.022
- VernacularTitle:热释光剂量仪测量体模植入125I粒子源的剂量学参数可行性
- Author:
Menglong ZHANG
;
Shanjun SONG
;
Weipeng WANG
;
Bin ZHAO
- Publication Type:Journal Article
- Keywords:
I-125 seed source;
Thermoluminescent dosimeter;
Phantom;
Dosimetry
- From:
Chinese Journal of Radiation Oncology
2014;23(2):165-168
- CountryChina
- Language:Chinese
-
Abstract:
Objective To study the dosimetry parameters of 125I seed source (type Sinko BT-125-1) with thermoluminescent dosimeter (TLD) in the phantom.Methods The new type of phantom was modified to suit to measurement of a common type of 125I seed source.The AAPM TG43 protocol recommended measurements of dose-rate constant (Λ),radial dose function (gL (r)),and anisotropy function (F (r,θ)) have been performed in the phantom with TLD.Results The Λ was 0.928 cGyh-1 U-1.The gL(r) was determined at different radial distances r ranging from 1.0 to 10.0 cm with an interval of 1.0 cm ; and F (r,θ) at angles from 0° to 90° in 10° increments.The gL (r) of 125I seed source showed a difference of 9.6% at the most in comparison to the corresponding values of 125I seed source (type Amersham 6711).The difference in F(2 cm,θ) of 125I seed source and Amersham 6711 was up to 10.2% near the source end.With the phantom the combined standard uncertainty in the whole measurement was less than 6.0%.Conclusions The experimental results exhibit fairly small measurement uncertainties and good self-consistency.It's feasible to measure the dosimetry characters of permanent implant seeds in the modified phantom.