Accuracy of index of consciousness in monitoring depth of anesthesia with sevoflurane inhalation in children: a comparison with bispectral index
10.3760/cma.j.issn.0254-1416.2013.09.004
- VernacularTitle:意识指数监测小儿吸入七氟醚麻醉深度的准确性:与BIS的比较
- Author:
Qian LI
;
Li ZHANG
- Publication Type:Journal Article
- Keywords:
Monitoring,intraoperative;
Anesthetics,inhalation;
Electrical trace technique;
Children
- From:
Chinese Journal of Anesthesiology
2013;33(9):1047-1049
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the accuracy of the index of consciousness (IoC) in monitoring the depth of anesthesia with sevoflurane inhalation in children,by comparing it with the bispectral index (BIS).Methods Forty-two ASA physical status Ⅰ or Ⅱ children,aged 8-12 yr,scheduled for elective abdominal surgery under general anesthesia,were studied.Anesthesia was induced with sevoflurane inhalation.The patients were endotracheally intubated.The concentration of sevoflurane was then adjusted to maintain the end-tidal concentration of sevoflurane (CETSev) at 1.6%,2.1%,2.6%,3.1%,3.6%,4.1% and 4.6% for5 min.The IoC and BIS values were recorded simultaneously.Results When CETSev < 3.1%,CETSev was negatively correlated with IoC and BIS values,and the correlation coefficients between CETSev and IoC and BIS values were-0.927 (r1) and -0.869 (r2),respectively (P < 0.01).When 3.1% ≤ CETSev ≤ 4.6 %,CET Sev was negatively correlated with IoC and BIS values,and the correlation coefficients between CETSev and IoC and BIS values were-0.808 (r3,P< 0.01) and-0.430 (r4,P < 0.05),respectively.Comparison of the absolute value of correlation coefficients:there was no significant difference between r1,r2 and r3 (P > 0.05) ; r1,r2 and r3 were significantly larger than r4 (P < 0.01).Conclusion IoC can accurately reflect the depth of anesthesia with sevoflurane inhalation in children and when 3.1% ≤ CET Sev ≤ 4.6 %,the accuracy of IoC in monitoring the depth of anesthesia is superior to that of BIS.