A comparative study on the clinical performance of three fourth generation HIV diagnostic reagents
10.3760/cma.j.issn.1009-9158.2013.10.011
- VernacularTitle:三种不同技术原理的第4代HIV诊断试剂的临床检测性能比较研究
- Author:
Jinming OUYANG
;
Xiaoxu HAN
;
Yangtao JI
;
Yanan WANG
;
Hong SHANG
- Publication Type:Journal Article
- Keywords:
HIV infections;
HIV;
Electrochemistry;
Enzyme-linked immunosorbent assay;
Indicators and reagents;
Reagent kits,diagnostic;
Comparative study
- From:
Chinese Journal of Laboratory Medicine
2013;36(10):903-907
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare the performance of fourth generation HIV antigen/antibody combined detection reagents for HIV early infection samples,international HIV seroconversion panel samples and routine clinical screening samples.Methods Thirty seven early HIV infected samples from the followup gays in Shen Yang between 2009 and 2011,66 seroconversion panel samples from BBI company (U.S.A),NABI company(U.S.A) and NIBSC company(U.K) and 703 routine HIV screening samples in the first hospital of China medical university in October 2010 were collected.All kinds of samples were tested by three diagnostic reagents based on chemiluminescence assay (CLIA),electrochemiluminescence assay (ECLIA) and enzyme-linked immunosorbent assay (ELISA) respectively.The detection sensitivity and specificity of these assays were analyzed.Results For 59 early infected and seroconversion samples,the sensitivities of both ECLIA and CLIA reagent were 96.61% (95% CI 91.5%-100.0%),higher than that of the ELISA kit (95% CI 75.0%-92.9%) (x2 =5.341,P < 0.05),which is 83.93% ; Comparison among the three reagents for different subtypes of the antibody seroconversion samples showed that ECLIA had the highest sensitivity while CLIA was the lowest ; Detection sensitivity of the three reagents for the P24 antigen is CLIA > ECLIA > ELISA; With detection of 703 clinical routine screening samples,the specificities of three reagents were 100% (CLIA),99.86% (ECLIA) and 99.71% (ELISA) respectively.Conclusions For the sensitivity of the fourth HIV diagnostic reagents CLIA and ECLIA are better than ELISA.The former two reagents are more suitable for identifying earlier HIV infection in clinic.