Method comparison between INNOVANCE D-Dimer and PLUS D-Dimer
10.3760/cma.j.issn.1009-9158.2013.07.016
- VernacularTitle:INNOVANCE试验和PLUS试验检测D-二聚体方法比较
- Author:
Ye GUO
;
Weiling SHOU
;
Wei WU
;
Qian CHEN
;
Xiaopei HAO
;
Wei CUI
- Publication Type:Journal Article
- Keywords:
D-Dimer;
Laboratory techniques and procedures;
Reagent kits,diagnostic
- From:
Chinese Journal of Laboratory Medicine
2013;36(7):638-642
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the clinical performance of INNOVANCE D-Dimer,and provide information for clinical application.Methods 402 cases of sodium citrate anticoagulant blood were tested with INNOVANCE assay and PLUS assay on CA7000 analyzer to measure plasma D-Dimer levels.VIDAS-30 immunology analyzer was also used to validate the two assays.4 patients with elevated D-Dimer were monitored continuously during 5 days using INNOVANCE assay and PLUS assay respectively,then the consistency of trend between 2 assays was analyzed.Plasma specimens added with hemoglobin,bilimbin and triglyceride were used to verify the anti-interference capability of INNOVANCE D-Dimer assay.Results In 402 specimens,the result ranges of INNOVANCE D-Dimer and PLUS D-Dimer were [2.15 (0.33,8.63)]mg/L FEU and [325.50 (123.75,974.00)] μ,g/L DDU,respectively.The consistency between two assays was poor (Z =-17.375,P =0.000),especially the results in the range of PLUS D-Dimer (201-300) μg/L DDU and (301-400) μg/L DDU,the coincidence rates were only 25% and 15%,respectively; the coincidence rate was up to 85% during PLUS D-Dimer (500-600) μg/L DDU; the coincidence rate was close to 100% when PLUS D-Dimer over 700 μg/L DDU.Totally 47 of 402 cases were unmatched between two assays.Verified by VIDAS 30,83.0% (39/47) was false negative for PLUS assay,4.3% (2/47) was false negative for INNOVANCE assay,12.7% (6/47) was false positive for PLUS assay.There were 5 false positives and 39 false negative for PLUS assay,totally 45 cases; Two false negative for INNOVANCE assay.Four patients with elevated D-Dimer were monitored and the results showed similar trend between 2 assays.For INNOVANCE assay,the capacity of anti-interference to free bilirubin,unconjugated bilirubin,hemoglobin,and triglyceride was up to 217 μmol/L,337 μmol/L,41.04 g/L,18.35 mmol/L,respectively.Conclusions INNOVANCE assay can markedly reduce false negative results of D-Dimer compared with PLUS assay.INNOVANCE D-Dimer has good performance on anti-interference to jaundice,hemolysis and lipemia samples.