The sensitivity of patient-specific VMAT QA to MLC positioning errors
10.3760/cma.j.issn.0254-5098.2013.04.013
- VernacularTitle:容积调强旋转放疗的计划验证通过率对多叶准直器位置误差的灵敏度
- Author:
Qingxin WANG
;
Jianrong DAI
;
Ke ZHANG
;
Xin YANG
- Publication Type:Journal Article
- Keywords:
Volumetric-modulated arc therapy VMAT);
Patient-specific QA;
Multileaf collimator (MLC);
Positioning errors;
Passing rate
- From:
Chinese Journal of Radiological Medicine and Protection
2013;33(4):388-391
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the sensitivity of patient-specific volumetric-modulated arc therapy (VMAT) quality assurance (QA) to minor multileaf collimator (M LC) positioning errors.Methods Systematic multileaf collimator (MLC) positioning errors (+0.5 mm,+ 1 mm and +2 mm) were introduced into the clinical VMAT patient plans with 2 types of MLC positioning errors:systematic MLC gap width errors and systematic MLC shift errors for 6 cases,including 3 cases with prostatic cancer and 3 cases with nasopharyngeal cancer.The planar dose distributions of the original and modified plans were measured using ArcCheck array.The coincidence between the measured results and the calculated results was evaluated using both absolute distance-to-agreement (AD-DTA) analysis with 3%/3 mm and 2%/2 mm criteria.Results The average passing rate of the 6 original VMAT plans was 96.0% with the ADDTA criteria of 3%/3 mm which was commonly adopted in clinical practice.For the MLC gap width errors of + 1 mm,+2 mm,and-2 mm and the MLC shift errors of 2 mm,the drop levels in average passing rate with the AD-DTA criteria of 3%/3 mm were 8.8%,15.5%,6.1% and 7.9%,respectively.The + 2 mm MLC positioning errors and + 1 mm MLC gap width errors could be detected by the patient-specific VMAT QA procedure.The AD-DTA criteria of 2%/2 mm was more sensitive compared with the criteria of 3%/3 mm.Conclusions Patient-specific VMAT QA is not sensitive enough to detect the systematic MLC positioning errors within 1 mm.Additional MLC QA is needed to guarantee the accuracy of VMAT delivery.