An investigation of a coincidence detection system for an all-digital small animal PET scanner
10.3760/cma.j.issn.2095-2848.2013.03.016
- VernacularTitle:全数字化小动物PET符合探测系统的设计与研究
- Author:
Wei LIU
;
Yuanbao CHEN
;
Anwen LONG
;
Xin CHEN
;
Zhongyi WU
;
Yongxue ZHANG
;
Qingguo XIE
- Publication Type:Journal Article
- Keywords:
Tomography,emission-computed;
Animal;
Software design
- From:
Chinese Journal of Nuclear Medicine and Molecular Imaging
2013;(3):223-225
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate and design a coincidence detection system for an all-digital small animal PET scanner and evaluate its preliminary performance properties.Methods This coincidence module adopted a coincidence identification mode based on singles data in list-mode.Using digital signal processing technology,energy calibration,crystal identification,timing alignment and coincidence events extraction were performed on singles data in list-mode.The obtained data could be used for image reconstruction.Results The 13 × 13 crystal array was well recognized by the position histogram of one lutetium yttrium orthosilicate (LYSO) crystal block.In the coincidence timing histogram of the micro-Derenzo phantom,1.36 ns full width at half maximum was obtained.The rods with a diameter of 1.2 mm were clearly displayed in the reconstructed image of the micro-Derenzo phantom.Conclusion The coincidence module can provide satisfactory performance to meet the design needs of an all-digital small animal PET scanner.