Design and application of accurately measured aseptic drainage device
10.3969/j.issn.1671-8283.2013.02.025
- VernacularTitle:准确计量无菌引流装置的设计与使用*
- Author:
Xuemei ZHENG
;
Lina QIAO
;
Xianzhen JIN
;
Chunyan LIAO
- Publication Type:Journal Article
- Keywords:
drainage device;
drainage liquid;
nursing
- From:
Modern Clinical Nursing
2013;(2):84-86
- CountryChina
- Language:Chinese
-
Abstract:
Objective To design a kind of aseptic drainage device with safe practice and accurate measurement for clinical nurses. Methods The experiment group included 63 cases using drainage device with accurate measurement,and control group included 63 cases using disposal drainage bag with scale of visual measurement.The accuracy of drainage liquid measurement was compared between two groups,and bacterial culture was made for the drainage liquid of two groups.Results The measurement error of experiment group and control group was(3.31±1.8)mL and(56.0±5.8)mL,respectively.The difference was significant(t=-4.593,P<0.01).The bacterial culture results showed no obvious difference between two groups(positive rate were 15.9%vs 17.5%,χ2=0.057,P>0.05). Conclusion The accurately measured aseptic drainage device is convenient for nurses to practice in clinics,and its application can reduce workload of nurses and decrease the occurrence of pollution.