Evaluation for clinical application of UniCel DxH 800 hematology analyzer
10.3760/cma.j.issn.1009-9158.2012.07.014
- VernacularTitle:UniCel DxH 800血细胞分析仪临床应用评价
- Author:
Chengwei PU
;
Ke SHANG
;
Jianzhong WANG
;
Xuekai LIU
;
Ying XING
;
Xiaoling TANG
;
Hui GAO
;
Ning DONG
;
Huiqing SU
;
Xiufen MING
- Publication Type:Journal Article
- Keywords:
Blood cell count;
Autoanalysis;
Diagnostic equipment
- From:
Chinese Journal of Laboratory Medicine
2012;35(7):643-646
- CountryChina
- Language:Chinese
-
Abstract:
ObjectiveTo evaluate the improvement on test performance of UniCel DxH 800 automated hematology analyzer for complete blood count (CBC) by detecting its performance indicators and comparing the differences of the results with LH 750 hematology analyzer and ADVIA 2120 hematology analyzer.MethodsThe precision,carryover and linearity of UniCel DxH 800 in measurement of CBC were evaluated by using fresh blood samples and instrument quality control of products.To evaluate the accuracy of leukocyte differential count and reticulocyte count with the microscopic method as the “gold standard”.To calculate the bias and correlation between the results measured by LH 750,ADVIA 2120 and UniCel DxH 800 hematology analyzers and compare these three instruments on the validity of the alarm in abnormal cells.ResultsIntra-precision:The coefficient of variation (CV) of the results of RBC,Hb and MCV were less than 0.5%,the CV of WBC and PLT results were less than 1.5%.Inter-precision:the CV of the parameters above were less than 2.5%.The carryover rate of WBC,RBC,Hb,MCV and PLT were less than 0.51%.In the concentration range covered by clinical samples,the correlation coefficients between the measured values and theoretical value in testing WBC,RBC,Hb and PLT were greater than 0.999 ( P <0.01 ).The measurement results of WBC,RBC,Hb,MCV and PLT hy UniCel DxH 800,ADV1A 2120 and LH 750 hematology analyzers have good correlation (r > 0.973,P < 0.01 ).Correlation of reticulocyte count between the UniCel DxH 800 hematology analyzer and microscolpic method was significant (r =0.920,P <0.01 ).Correlation of leukocyte differential count about the grauulocytes, lymphocytes and eosinophils between the UniCel DxH 800 hematology analyzer and microscopic method was good (r =0.914,0.900 and 0.725,P <0.01 ),followed by monocytes ( r =0.612,P <0.01 ),which were better than the LH 750 with similar detection principle.The UniCel DxH 800 hematology analyzer demonstrated higher sensitivity (96.6% ) for the alarm of abnormal cells and achieved a lower false-negative rate (2.5% ).Meanwhile,the sensitivity of the neutrophil nuclei left shift was higher (90.5% ) and the false-negative rate (5.0%) was lower.ConclusionsThe UniCel DxH 800 hematology analyzer for complete blood count shows advantages of high precision,low carryover rate and wide linear range.The results detected by the UniCel DxH 800 hematology analyzer have good correlation with the LH 750 and ADVIA 2120.