Evaluation of the short-time effect of radiofrequency ablation on kidney by contrast-enhanced ultrasonography and magnetic resonance imaging
10.3760/cma.j.issn.1004-4477.2012.05.024
- VernacularTitle:超声造影、磁共振评价肾射频消融短期效果的实验研究
- Author:
Fanhua XU
;
Bing HU
;
Bulin ZHANG
;
Shengli KUANG
- Publication Type:Journal Article
- Keywords:
Ultrasonography;
Microbubbles;
Kidney;
Cathecer ablation;
Magnetic resonance imaging
- From:
Chinese Journal of Ultrasonography
2012;21(5):439-441
- CountryChina
- Language:Chinese
-
Abstract:
ObjectiveTo evaluate the accuracy of the modalities in revealing the results by observing the radiofrequency ablated (RFA) lesions with contrast-enhanced ultrasonography (CEUS),magnetic resonance image (MRI) in comparison with histopathologic findings in normal rabbit kidney.Methods RFA was performed in the left kidneys of 10 rabbits.The ablated lesions were observed with conventional ultrasound(US),CEUS,MRI at 1 day,1 week after RFA.Then the rabbits subjected to postmortem and the kidney specimens were excised,stained and processed for conventional histologic study.The maximum lesion size as measured on US,CEUS and MRI was then correlated with the histopathologically determined lesion size.The utility of different imaging modalities were evaluated.ResultsIn a relatively short time (at 1 day and 1 week),CEUS and MRI may reveal the location,shape and extent of ablated lesion.And the lesion sizes with CEUS,MRI and kidney specimens were well correlated (P>0.05).ConclusionsBoth CEUS and MRI are suitable for the accurate evaluation of the ablated lesions in the short time.CEUS can reveal the variation of flow perfusion in the ablated lesion dynamicly and in real time.