Quantifying and correcting underestimation of cardiac Doppler blood flow velocities using dual PW/DTI technique
10.3760/cma.j.issn.1004-4477.2011.12.001
- VernacularTitle:应用脉冲波/组织多普勒双模式技术定量测定和矫正心脏多普勒血流测量低估值
- Author:
Zhen WANG
;
Tiesheng CAO
;
Lijun YUAN
;
Yong YANG
;
Jie LIU
;
Lianbi ZHAO
;
Yunyou DUAN
- Publication Type:Journal Article
- Keywords:
Echocardiography;
Blood flow velocity;
Heart valves;
Pulse wave Doppler;
Tissue Doppler imaging
- From:
Chinese Journal of Ultrasonography
2011;20(12):1013-1016
- CountryChina
- Language:Chinese
-
Abstract:
ObjectiveTo quantify the underestimation of cardiac blood flow velocities measured by Doppler echocardiography and to explore a method for correction using a new explored dual PW/TDI technique.MethodsThe dual PW/DTI mode was used to simultaneously record the aortic,mitral,tricuspid and pulmonary valvular blood flow velocity and the adjacent valvular annulus velocity in forty healthy volunteers,then the underestimations of the flow velocities were calculated.ResultsThe true blood flow velocity relative to the valvular annulus could be obtained by dual PW/TDI technique.Conventional Doppler echocardiographic measurements significantly underestimated the true velocity ( P < 0.001).The actual aortic blood flow velocity had a significant underestimation of (8.5 ± 1.2) %,the actual flow velocity of the pulmonary artery had a significant underestimation of (6.6 ± 1.1) %,the underestimations of E and A wave of mitralvalve's were (13.4 ± 1.7)% and (16.7± 3.4)%,the underestimations of E and A wave of tricuspid valve' s were ( 18.7 ± 1.9) % and (26.0 ± 16.1 ) %.ConclusionsThe cardiac blood flow velocity measured by Doppler ultrasound has been significantly underestimated and this underestimation can bequantified and corrected using dual PW/DTI technique based on the principle of motion relativity.