Dose comparison of different scan projections of Implagraphy cone beam computed tomography for dental maxillofacial use
10.3760/cma.j.issn.0254-5098.2012.05.025
- VernacularTitle:Implagraphy牙颌面锥形束CT不同扫描体位的辐射剂量对比研究
- Author:
Dong FANG
;
Xianshun YUAN
;
Dongsheng ZHANG
- Publication Type:Journal Article
- Keywords:
Cone beam computed tomography;
Absorbed dosage;
Equivalent dosage;
Effective dosage
- From:
Chinese Journal of Radiological Medicine and Protection
2012;32(5):535-538
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the subject's absorbed dose,equivalent dose and effective dose.Methods The CBCT unit was Implagraphy and three scan projections were selected such as mandible,maxilla and temporamandibular joint (TMJ).Thermoluminescent dosimeter tubes were used to record the absorbed dose at special positions in the head and neck region of an adult skull and tissue-equivalent phantom.16 interested organs included pituitary,lens,parotid glands,submandibular glands,sublingual glands、diploe,spongy bone of the chin and cervical vertebra,skins of cheeks and nuchal region,thyroid and esophagus.The absorbed dose was measured in these organs,and then the effective dose(E1990,E2007)were calculated according to different ICRP tissue weighting factors.Results The absorbed dose of mandible,maxilla and TMJ scan varied from (0.99 ±0.09) to (12.85 ±0.09)mGy,(0.93 ±0.01) to (13.07 ±0.02) mGy and (0.68 ±0.01) to (10.18 ± 0.04)mGy.There was significant difference among the three scan projections (F =19.61-30992.27,p < 0.05).The equivalent doses of lens and skin were (1.11± 0.07)-(5.76 ± 0.06) mSv and (6.96 ± 0.06)-(10.64 ± 0.07) mSv.There was significant difference among the three scan projections(F =4473.02,9385.50,P <0.05).The effective dose(E1990,E2oo7) was [(191.35±1.53),(325.17 ±2.58) μSv] for mandible scan,[(106.62 ±2.17),(226.28 ±2.81)μSv] for maxilla scan,[(104.21 ± 1.02),(142.36 ± 1.90) μSv]for TMJ scan,respectively.Conclusions The valid measurement should be taken to reduce the subject' s dose such as a careful history and clinical examination before the performance of CBCT,the latest risk/benefit assessment,precise scan position,the shielding of thyroid as well as brain and the smaller volume size as well.