Analysis of international intercomparisons results organized by Japan for integrating 222Rn-220Rn detectors
10.3760/cma.j.issn.0254-5098.2012.05.017
- VernacularTitle:222Rn-220Rn累积探测器参加日本组织的国际比对结果分析
- Author:
Yunyun WU
;
Hongxing CUI
;
Qingzhao ZHANG
;
Bing SHANG
;
Xu SU
- Publication Type:Journal Article
- Keywords:
222Rn;
220Rn;
Integrating radon/thoron detectors;
International intercomparison
- From:
Chinese Journal of Radiological Medicine and Protection
2012;32(5):514-517
- CountryChina
- Language:Chinese
-
Abstract:
Objective To guarantee the quality of measurenents with the radon-thoron discriminative detectors of our laboratory.Methods LD-P radon-thoron discriminative detector participated in the international intercomparison for integrating radon/thoron detectors organized by National Institute of Radiological Science (NIRS,Japan).Detectors were sent to NIRS for exposure.Radon intercomparison was conducted with radon chamber providing three levels of exposure:low,medium and high levels.Thoron intercomparison was carried out at thoron chamber,which also provided three levels of exposure:low,medium and high levels.Detectors were posted back to our laboratory for etching and analysis after exposure.Then the measured values were submitted to NIRS.Finally the reference values were informed of us.Results The relative percent difference (RPD) between the measured value and the reference value for radon was-13.8%,-14.4% and-17.1% at low,medium and high levels respectively,and that of thoron were-14.4%,8.9% and-3.2% at three levels respectively.Conclusions Both radon and thoron measurement of our detectors rank asCategory Ⅰ in the 4th international intercomparisons for integrating radon/thoron detectors with the NIRS radon/thoron chambers.