Monte Carlo calculation of the impact on dose distribution by width of energy spectrum and angular distribution of electron beam
10.3760/cma.j.issn.1004-4221.2012.01.022
- VernacularTitle:电子束能谱宽度及角分布对蒙特卡罗方法模拟计算剂量分布影响
- Author:
Shengwei KANG
;
Pei WANG
;
Jinyi LANG
;
Qing HOU
;
Dake WU
- Publication Type:Journal Article
- Keywords:
Radiotherapy,electron;
Monte Carlo;
Dosimetry
- From:
Chinese Journal of Radiation Oncology
2012;21(1):80-82
- CountryChina
- Language:Chinese
-
Abstract:
ObjectiveTo study of the impact on dose distribution by electron width of energy spectrum and angular distribution using Monte Carlo simulation.MethodsThe simulated electron energy spectrum and angular distributions was as the input parameters and percentage depth dose (PDD) and offaxis curves were simulated by a modified PENELOPE code package. Results PDD and off-axis dose distribution curves are almost the same and are not sensitive to energy spectrum width and angular distribution with the exception of energy spectrum width of 2.5 MeV with obviously different curves.ConclusionsIn the situation of clinical treatment,spectrum and angular distributions can be ignored when their width are not very large.It is helpful to save about 9% time by using monoenergy beams in treatment planning system development.