Optimized imaging quality and radiation dose for coronary artery angiography using 128-slice,dual-source Flash Spiral CT under the natural heart rate
10.3760/cma.j.issn.1005-1201.2011.05.014
- VernacularTitle:自然心率下128层双源Flash Spiral CT冠状动脉成像质量及辐射剂量的研究
- Author:
Yuejun XUE
;
Nong QIAN
;
Yanhui SHAO
;
Changjie PAN
;
Weiliang RONG
;
Yiqun XU
;
Zhiwei TAO
- Publication Type:Journal Article
- Keywords:
Radiation doses;
Tomography spiral computed;
Coronary vessels;
Heart rate
- From:
Chinese Journal of Radiology
2011;45(5):481-485
- CountryChina
- Language:Chinese
-
Abstract:
Objective To compare the quality and radiation doses of coronary artery angiography under the natural heart rate condition between Flash spiral heart mode and prospective electrocardiogramtriggering sequence mode using dual-source,in order to choose personalized low doses of coronary artery scanning mode.Methods Sixty patients who underwent coronary angiography(CTA)on a 128-slice,dualsource CT scanner were divided into 2 group i.e,group A(27cases)and group B(33 cases).Flash spiral heart scan mode was employed for group A.Inclusion criteria included:heart rate<65 bpm.regular sinus rhythm,heart rate fluctuation less than ±5 bpm.Date acquisition was set at 60% of the R-R interval.Prospective electrocardiogram-triggering sequence scan mode(SAS)was performod for group B.Inclusion criteria included:(1)heart rate≥65 bpm,(2)arrhythmias,premature beat,fibrillation atrial.Exclusion criteria included:bad holding breath.Date acquisition(1)At low heart rate(≤75 bpm),date acquisition was set at 60%-80%of the R-R interval.(2)At high heart rate(>75 bpm),date acquisition was set at 30%-50%of the R-R interval. (3)At the arrhythmias,premature beat,fibrillation atrial,date acquisition was set at 20%-90%of the R-R interval.In both gronps,patients with a BMI≥25.0kg/m2 were examined with a tube voltage of 120 kV.while the other patients with a BMI<25.0 kg/m2 were examined with a tube voltage of 100 kV.The BMl was(24.6±1.0)kg/m2 in group A,while that was (24.6±0.9)kg/m2 in group B.In both groups,all images were transferred to the workstation for further processing and analysis.The imaging quality of coronary artery segments and the radiation dose were compared with t test.Results A total of 336 coronary artery segments were evaluated in group A and 412 segments were evaluated in group B.The imaging quality of coronary artery segments were scored.Excellent or good was achieved in 98.2%(330 of 336)artery segments in group A,and that was 98.1%(404 of 412)in group B.There was no statistical difference in imaging quality between the two groups(t=0.513,P=0.608).The average effective dose was(0.74±0.29)mSv in group A,whereas that was(3.67±1.37)mSv in group B.There was a significant difference between the two groups(t=-10.858,P=0.000).Conclusions The personalized low doses coronary artery scanning mode can substantially reduce radiation damage while preserving good imaging quality.