Commissioning of intensity - modulated radiation therapy system with independent collimator
10.3760/cma.j.issn.1004-4221.2011.02.022
- VernacularTitle:独立准直器调强放疗系统的验收测试
- Author:
Ke ZHANG
;
Jianrong DAI
;
Li CHEN
;
Yuheng LI
;
Liang LI
- Publication Type:Journal Article
- Keywords:
Intensity-modulated radiotherapy system;
Independent collimator;
Commissioning
- From:
Chinese Journal of Radiation Oncology
2011;20(2):164-167
- CountryChina
- Language:Chinese
-
Abstract:
Objective To perform an acceptance test for the IMRT system with independent collimator. Methods An ion chamber dosimeter were used to measure the startup characteristics of the accelerator and the absolute dose at isocenter and given characteristic points for three clinical cases ( a lower nasopharyngeal carcinoma, a lung cancer and a cervical cancer). The characteristic points represented the organs at risk or the target. A Mapeheck2 was used to measure dose maps of basic test fields and the treatment fields for the clinical cases. The basic test fields were as follows: 1 ). Symmetric fields in size of 2 cm ×2 cm, 5 cm ×5 cm, 10 cm× 10 cm, 20 cm ×20 cm, 2 cm × 10 cm, 10 cm ×2 cm, 5 cm ×20 cm and 20 cm ×5 cm;2). Asymmetric fields in size of 2 cm ×2 cm (x1 =4 cm, y1 = 10 cm;x2 = -2 cm, y2 = -8cm) and 5 cm ×5 cm (x1 = -2 cm, y1 = -5 cm;x2 =7 cm, y2 = 10 cm) ;3) A 20 cm ×20 cm composite field composed of five 20 cm× 4 cm narrow bar fields side by side. Gamma Index was used to compare calculated and corresponding measured dose distributions. When the criterion was 3% dose difference or 3 mm distance-to-agreement, the pass rate was required to be more than 90%. Results The accuracy of machine output was better than 2% when machine monitor units increased to 4. Among all basic test fields and all the treatment fields of three clinical cases, the maximal absolute dose error was -3.67%, and only the composite test field and two treatment fields of the lower nasopharyngeal carcinoma case had a pass rate slightly less than 90%, which were 83.6%, 88. 3% and 89. 7% ,respectively. For the three clinical cases the treatment delivery times were 15, 14, and 27 minutes, respectively. Conclusions The overall commissioning results are acceptable, and the system can be used in clinic.