Rapid identification and susceptibility testing of positive blood culture caused by gram negative bacteria
10.3760/cma.j.issn.0254-5101.2011.03.006
- VernacularTitle:革兰阴性菌引起的血培养阳性标本的直接鉴定药敏试验
- Author:
Yueru TIAN
;
Feiyi RUAN
;
Hong LIU
;
Fuqi AI
;
Yimin MA
;
Ye JIANG
;
Xiaofei JIANG
- Publication Type:Journal Article
- Keywords:
Positive blood culture;
Joint use;
Rapid identification and susceptibility testing
- From:
Chinese Journal of Microbiology and Immunology
2011;(3):220-224
- CountryChina
- Language:Chinese
-
Abstract:
Objective To reduce the turnaround time for laboratory diagnosis of bacteremia, the feasibility of rapid identification and susceptibility testing using samples taken directly from positive blood culture bottles was evaluated. Methods The growth of microorganisms in blood culture bottles was screened by the BACTEC 9000 blood culture system. 65 positive blood culture bottles containing gram-negative bacteria were adopted to test. Culture fluid was injected into BD SST vacutainer and centrifuged to pellet blood cells. After collecting required McFarland units, they were cultured on Phoenix 100 NMIC/ID-4(identification-gram-negative bacteria and susceptibility testing) cards using 0.25 McF and 0.5 McF methods respectively. They were also evaluated by the standard method, involving subculture tests from positive blood culture bottles. Results 63 of 65 gram-negative bacteria (96. 9% ) were correctly identified with 0. 25 McF method. 59 of 65 gram-negative bacteria(90.8% ) were correctly identified with 0.5 McF method. For antimicrobial susceptibility testing, the 0.25 McF direct method had an agreement rate more than 94% , the 0.5 McF method was more than 85.7% and direct blood sample KB method was more than 93.8% compared to the standard method. But the overall minor error rate in susceptibility testing of direct blood sample KB method is higher than other methods. Conclusion Applying 0. 25 McF and 0. 5 McF rapid identification and susceptibility test was practical. During to possessing more prominent advantages, laboratory put the 0. 25 McF direct method into practice had a timely, remarkable significance.