The Comparison of the Fixation Strength of Locking and Non-locking Anterior Cervical Plate Systems
- VernacularTitle:颈椎前路带锁与非带锁钢板螺钉系统固定强度的比较
- Author:
Deyu CHEN
;
Qiang SHEN
;
Dianwen SONG
- Publication Type:Journal Article
- From:
Orthopedic Journal of China
2001;8(2):148-149
- CountryChina
- Language:Chinese
-
Abstract:
Objective: To compare the fixation strength of locking and non-locking anterior cervical plating initially and after fatigue.Methods:AO Cervical Spine Locking Plate(CSLP),Danek Orion plate and AcroMed Acroplate were fixed on lamb cervical spine.The fixation strength and pull-off strength of the screw-plate constructions was performed initially and after fatigue.Results:Locked CSLP and Orion constructions were more rigid than all unlocked unicortical systems initially and after cyclic loading.There was no significant difference in pull-off strength between the CSLP,the Orion,and the unicortical Acroplate.Conclusions:Locking mechanism significantly increased the strength of the tested screw-plate systems initially and after fatigue.