Determination of Selenium in Hair by Hydride Generation Atomic Absorption Spectrophotometry
- VernacularTitle:氢化物发生原子吸收分光光度法测定发硒含量
- Author:
Huiqi LIU
- Publication Type:Journal Article
- From:
Journal of Environment and Health
2001;18(2):115-116
- CountryChina
- Language:Chinese
-
Abstract:
The levels of selenium in hair were determined by hydride generation atomic absorption spectropho-tometryin this assay. The working conditions of the determination were controlled as follows,wavelength: 196.00 nm,working current of lamp:6 mA,wide of filtered spectrum:4.0 A, flow rate of gaseous argon: 120 ml/min. The averagecontents of selenium in hair samples reached 0. 0498±0. 031 mg/kg. The detection limit and coefficient of variation were0. 5 ng/ml and 5.8% respectively. The recovery rates were 92.0%~ 104. 0% with an average of 99. 5%. This methodshowed advantages ,such as simple operation,rapid,low cost and high accuracy. The obtained data were reliable.