Determination of Serum Concentrations of Propofol by RP- HPLC
- VernacularTitle:反相高效液相色谱法测定异丙酚血清浓度
- Author:
Shuomin JIANG
;
Zengshou WANG
;
Shengwei JIN
- Publication Type:Journal Article
- From:
China Pharmacy
2001;12(3):161-162
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE: A RP- HPLC method was established for determination of serum concentration of propofol in man METHODS: The mobile phase consisted of methanol- water( 68∶ 32v/v) The detection was carried out at wave length 258nm and flow rate 1ml/min with carbamazepine as internal standard RESULTS: The retention times of propofol and carbamazepine were 9 20 and 5 16 min, respectively The mean recovery of propofol was 99% The within- day and inter- day variations were all less than 10% Propofol and carbamazepine were seperated well The assay linearity was obtained in the range of 1~ 16μ g/ml in serum( r=0 9 994) CONCLUSION: The method is sensitive, simple and reliable for the determination of propofol concentration